Meter
The meter facility allows you to monitor sensors and adapters without actually
storing data. When DataMeter is working as a meter a small M appears in the
bottom left of the display. Plug sensor(s) into any socket and the display will show
sensor SI units and readings directly below.
•Battery or Power pack
There are icons to indicate battery and charging conditions which are explained
in the overview section at the front of this manual. Only use the mains power
pack/charger supplied (or the original LogIT SL power pack). If it does not have
LogIT on it, DON'T USE IT!
•Power save
To save battery power DataMeter switches itself off if no button has been
pressed for 2 minutes. However you can override this by selecting 'Meter stay on'
from the main menu - the M symbol on the screen will flash and DataMeter will
then stay on until you switch it off or start logging. If you are using DataMeter with
the power pack/charger then the display will automatically stay on until you
switch it off.
Tip A quicker way to make the meter stay on is to hold green button down for
more than 2 seconds when you switch DataMeter on.
Real Time Logging
I
n real time DataMeter acts as a 'live' interface between sensors and the
computer. The computer has total control over logging rates/time span and
DataMeter sends back the sensor calibration and live data (no data is stored in
DataMeter). As the computer has so much control, actual operation and function
of real time logging will very much depend on which computer and software you
have. However, here are the basics (it may also be useful to refer to the Real Time
logging set-up picture on page 6):
• Connect DataMeter to your computer using the link cable
• Load datalogging software on computer (also select
DataMeter as the interface type if required by your software)
• Set up logging rate and duration of experiment as required.
• Insert sensors required and check readings on DataMeter display
• Start logging with software (DO NOT press DataMeter's
green button). Display should say 'Computer Link on'.
• When required select stop logging on the computer
Tip The speed of real time datalogging is limited by the relatively slow serial link
between DataMeter and Computer. If you need fast datalogging use DataMeter
in remote AutoLog or Programmed logging mode and then download to your
computer.
10
Downloading the stored data
Having logged data in DataMeter using AutoLog, Snapshot or Programmed
logging you need to retrieve and use it.
There are basically two ways of downloading or 'exporting' data.
1 Dedicated Datalogging software
LogIT brand software, and most third party datalogging software supports
DataMeter (see 'Software' in the reference section). With datalogging software
all you have to do is connect DataMeter to your computer using the link cable
and use an appropriate menu to load data. You may be given a choice from 4
sets of results to load because DataMeter can store up to 4 experiments data.
You do NOT need to press DataMeter buttons - it will link automatically.
LOAD
SET 1
SET 2
SET 3
SET 4
serial
socket
2 Sending Data direct to the computer with comms software
If you do not have datalogging software or if you just want to transfer the raw
data into a spreadsheet etc you may be able to receive data with a
communications or 'works' program. Eg:
IBM PC compatible
Windows Terminal, Microsoft Works
Apple Macintosh
Apple Terminal, ClarisWorks, MS Works,
Acorn Archimedes
ArcComm 2, Hearsay
BBC Master
Built-in terminal program
Psion/Acorn Pkt Bk
Serial/A-Link software (built into link)
The comms software should be set to 9600 Baud, 8 bit, no parity, 1 stop bit, RTS/CTS
handshaking. Then select 'Send Data File' from the DataMeter menu & choose
CSV (comma separated value) or TAB (tab sep. value) for spreadsheets, or SID for
datalogging software. The last data set will be sent. For details see README on
system disc.
•Printing data direct to a printer
You can print a table of ASCII text data direct to a serial printer using 'Send Data
File'. The printer should be set to same settings as above.