Agrowtek SXW Manual de instruções - Página 9

Procurar online ou descarregar pdf Manual de instruções para Registadores de dados Agrowtek SXW. Agrowtek SXW 11 páginas. Weather station sensor & data logger

For example: to set the temperature to a calibrated value of 25°C, write the value "2500. "
Analog Calibration (±1 calibration bit = ±0.005mA adjustment)
Analog output calibration sends a positive or negative o set to the respective output channel's digital to
analog converter (DAC.) The DAC has a resolution of 0.005mA/bit.
For example: to shift the analog output up by 0.1mA, set the analog o est value to +20. ( 0.1 / 0.005 = 20)

MODBUS Holding Registers

Parameter
Address
Serial#
DOM
HW Version
FW Version
Toggle Units
Sensor Reading,
Integer
Sensor Reading,
Float
Calibration Input,
O set (Zero)
Calibration Input,
Analog Output
A request to read or write a register that is not available will return an illegal address exception.
9
Description
Range
Device Slave Address
1 - 247
Device Serial Number
ASCII
Date of Manufacture
ASCII
Hardware Version
ASCII
Firmware Version
ASCII
Toggle sensor units
1 - 4
Temperature
-2000 - 6000 (-20 - 60°C)
Humidity
0 - 1000 (0 - 100%)
Light
0 - 1000 W/m2
CO2
0 - 10,000ppm
Speed
0 - 125mph
Direction
0 - 359°
N/A
Barometric Pressure
Rain
0 - 1
Temperature
-20.00- 60.00 °C
Humidity
0 - 100.0 %
Light
0 - 1000 W/m2
CO2
0 - 10,000ppm
Speed
0 - 125mph
Direction
0 - 359°
Barometric Pressure
N/A
Temperature
Humidity
Light
CO2
See integer ranges above.
Speed
Direction
Barometric Pressure
Temperature
Humidity
-255 - 255 (bits)
Light
CO2
© Agrowtek Inc. | www.agrowtek.com | Technology to Help You Grow™
Type
Access
8 bit
R/W
8 char
R
8 char
R
8 char
R
8 char
R
16 bit, unsigned
W
16 bit, signed
R
32 bit, oating pt R
16 bit, signed
W
16 bit, signed
W
Address
40001
40004
40008
40012
40016
41002
40101
40102
40103
40104
40105
40106
40107
40108
40201
40203
40205
40207
40208
40211
40213
41101
41102
41103
41104
41105
41106
41107
41301
41302
41303
41304