How to make calibration
*At the time when a measurement is taken, pls be sure to perform it with the Sensor Cap on.
[Zero calibration]
1. Please turn on power and wait for 20 seconds
2. Please press the MODE button to switch to the Salinity Measurement Mode.
3. A long press of the MODE button will enable the display panel to blink. At that time you take your hand off
the MODE button and wait, when the zero calibration is carried out and the mode will be automatically
returned to the Salinity Measurement Mode.
[Solution calibration]
1. After the zero calibration is completed, please wait for 20 seconds steeping the DMT-10 in the standard solution.
2. A long press of the CAL button will enable the display panel to blink. At that time you take your hand off the CAL button
and wait, when the 30 ppt calibration is carried out and the mode will be automatically returned to
the Salinity Measurement Mode.
Caution: It is required that the sensor be completely immersed in the solution.