Rotor balancing.
Balancing of tubes for Micro and Fixed angle rotors
Swing out rotor, buckets. Balancing

Very Important

Tare (zero) your scales first
Weigh a tube , note its
weight Weigh another and
match its weight to within 0.1
gram
Place all your required tubes in all 4 buckets.
Equal amount per bucket.
Weigh each bucket containing tubes
Match other 3 buckets, with tubes to within 1
gram of each other
For BRK5506 rotor use the adaptor
4