FM 10-67-1 APPENDIX E
Depress the switch button on the instrument pack and read the calibration
meter. If the meter reads "0," the water detector is calibrated as shown in
Figure E-6. If the meter reads any increment other than "0," you must
adjust the internal calibration screw until the water detector is "zeroed-
out."
To adjust the internal calibration screw, remove the outer screw on the
side of the calibration meter housing using the jeweler's screwdriver.
Insert the jeweler's screwdriver into the housing and turn the internal
adjustment screw. Depress the switch button on the instrument pack as
you turn the screw until the meter lever reads "0." Replace the outer
screw and the water detector is calibrated. See Figure E-7.
Remove the calibration pad from the test area, using tweezers, and place
it back in the kit.