Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions
for common problems encountered with the Alnor AXD610/A
PVM610. If your symptom is not listed, or if none of the solutions solves
your problem, please contact TSI.
Table 5-1: Troubleshooting the Alnor AXD610/A
Symptom
No Display
Velocity reading
fluctuates unstable
Instrument Error
message appears
The pressure sensor is protected from damage up to 7 psi (48 kPa or
360 mmHg). At higher pressure it can burst!
Possible Causes
Unit not turned on
Low or dead batteries
Dirty battery contacts
Fluctuating flow
Fault in instrument
WARNING
IRFLOW
PVM610
IRFLOW
Corrective Action
Switch unit on.
Replace batteries.
Clean the battery
contacts.
Reposition probe in
less-turbulent flow.
Factory service required
on instrument.
!
9