2M Technology 2MTHFR-2M Manual do utilizador - Página 6
Procurar online ou descarregar pdf Manual do utilizador para Terminais tácteis 2M Technology 2MTHFR-2M. 2M Technology 2MTHFR-2M 8 páginas. Thermo detector
Também para 2M Technology 2MTHFR-2M: Manual rápido (17 páginas), Manual rápido (19 páginas)
Note: need set ID number only when adding optional network(LAN) function.
Installation Note
1. Install device in windless indoor, Cold wind could interfere thermo detection, we need turn up the ambient tem-
perature compensation factor (5 °C each time)
2. Sunshine could interfere thermo detection.
Thermo Detection Instruction
1. Normal use environment temperature: 0°C- 35°C; when using at a temperature of about -10°C, the ambient tem-
perature compensation value needs to be increased each time ( Adjust 5°C), until the error between the measured
value and the forehead gun is within ± 0.3, and then measure. If multiple body temperature measurements occur
Low, you can also use this method to calibrate.
2. Measurement method: forehead temperature measurement or wrist temperature measurement in same mode .
3. The best measurement distance is 1-10cm.
4. Measurement time is 0.5s, please move forehead or wrist after detection to avoid repetitive measurement or
non-responsive situation.
5. Environment could affect the thermo detection, the measurement accuracy is ±0.3°C.
Specifications
Product Weight
Weight
Working Temperature
Operating Humidity
Power Consumption
External Power Supply
Working Voltage
Temperature Measurement Speed
Temperature Measurement Accuracy
Temperature Measurement Distance
Dimensions
6
Weight 0.6kg
less than 8W
-10 ~ 45 °C
95%, non-condensing
Power consumption
AC100~240V, 50/60Hz
12V
Less than 1 second
± 0.3°C
1+20 cm
164 x 110 x 32mm (excluding
bracket and power adapter)
www.2mtechnology.net