SMA-plug
Connect the SMA connector of the fiber to the SMA-plug of the Light Source.
I/O connector
The signals on the I/O connector are described below.
Pin configuration
Pin
Name
10
GND
11
Disable
Input
Below the electrical characteristics for the input signals
Electrical characteristics
Name
Disable

2.2 12VDC power plug

The power supply specifications are mentioned in the table below
Name
Voltage
Power

2.3 Settings for disable

With the disable function that is present on the I/O connector the user can disable the optical
output of the light source by setting this signal to high-level.
To use the disable function the ON-OFF switch must be set in the ON mode.
When the light source is used in combination with an AvaSpec spectrometer, interface cable IC-
DB26-2 and AvaSoft, the optical output can be changed by AvaSoft. It is also possible to disable and
enable the light source by AvaSoft.
Revision: 2-00
© Avantes BV
Description
Ground
Input, used to disable/shut down the light source
Low level detection
< 1.0V
Minimum
10V
3Watt
www.avantes.com
High level detection
> 2.0V
Typical
12V
Maximum voltage
12.0V
Maximum
15V