Recommended Use Methods
1.Recommendation that the PATIENT relax as much as possible and not talk
during the measurement PROCEDURE
2.Recommendation that 5 min should elapse before the first reading is taken
3.Any reading can be affected by the measurement site, the position of the
PATIENT, exercise, or the PATIENT'S physiologic condition
4.Performance of the AUTOMATED SPHYGMOMANOMETER can be
affected by extremes of temperature, humidity and altitude
5.To stop the inflation or measurement, push the START/STOP button. The
monitor will stop inflating, start deflating, and will turn off.
6.After the monitor has detected your blood pressure and pulse rate, the cuff
automatically deflates. Your blood pressure and pulse rate are displayed.
7.The monitor will automatically turn off after one minute.
5.5. Irregular Heartbeat Detector
This symbol - indicates that certain pulse irregularities were detected during
the measurement.
In this case, the result may deviate from your normal basal blood pressure –
repeat the measurement.
Information for the doctor on frequent appearance of the Irregular Heartbeat
Symbol.
This instrument is an oscillometric blood pressure monitor device that also
analyzes pulse frequency during measurement. The instrument is clinically
tested.
Picture-07
If pulse irregularities occur during measurement, the irregular heartbeat
symbol is displayed after the measurement. If the symbol appears more
frequently (e.g. several times per week on measurements performed daily) or
if it suddenly appears more often than usual, we recommend the patient to
seek medical advice. The instrument does not replace a cardiac examination,
but serves to detect pulse irregularities at an early stage.
5.6. Error Indicates
The following symbol will appear on the display when measuring abnormal
SYMBOL
CAUSE
No display
Weak battery or improper
appears
placement
Er 1
Sensor abnormal
Monitor could not detect pulse
Er 2
wave or cannot calculate the
blood pressure data
Measurement result is abnormal
Er 3
(SYS≦45mmHg, DIA≦24mmHg)
Too loose cuff or air leakage
(Cannot inflate to 30mmHg within
Er 4
15s)
Er 5
The air tube is crimped
Er 6
The sensor is sensing great
fluctuation in the pressure
The pressure that the sensor
Er 7
sensing is over the limit
The demarcation is incorrect or the
Er 8
device has not been demarcated
CORRECTION
Replace both batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Check if the pump is working or not. If it
is working, then the problem is sensor
abnormal. Please send it to the local
distributor.
Check if the air releasing is too slow or not.
If it is too slow, please check if there is any
dust in the tube plug of the cuff and the cuff
port in the device. If yes, please clean and
start the measurement again. If no, please
send the device back to the local distributor.
Occasionally-measure for one more time/
Always - send it to local distributor
Tie the cuff correctly and make sure the air
plug is properly inserted in the unit
Correct it and make the measurement again
Please keep quiet and don't move
Please send back to the local distributor
Please send back to the local distributor