Bilanciai EV7-S Manual - Página 3
Procurar online ou descarregar pdf Manual para Terminais tácteis Bilanciai EV7-S. Bilanciai EV7-S 6 páginas.
10.2 Error messages
Only one code is displayed with the errors. To delete it, reset the
terminal.
Problem
-01-
Converter failure
-02-
EEPROM error
-04-
RAM checksum
error
-05-
EPROM
checksum error
Cause
Scale connector
disconnected or
broken
Conversion
board failure
Voltage drop
during an
incorrect
operation
CPU board fault
1-116
EV7 -S
Remedy
Switch the terminal off and
make sure that the connector
is correctly connected. If
necessary, disconnect it,
re-connect it and then power
the terminal again.
Contact the Service Centre.
Reset the instrument. If the
error persists, contact the
Service Centre for possible
replacement of the CPU
board.
Reset the instrument; If the
error persists, contact the
Customer Service for possible
replacement of the CPU
board.