DH Instruments PG7000 Instruções de atualização - Página 3
Procurar online ou descarregar pdf Instruções de atualização para Acessórios DH Instruments PG7000. DH Instruments PG7000 4 páginas. Flash with eeprom to flash without eeprom, non-flash to flash without eeprom
Também para DH Instruments PG7000: Instruções de instalação (2 páginas)
PG7000 ™ Upgrade Instructions Flash with EEPROM to Flash without EEPROM,
PG FLASH MEMORY
ADAPTOR (SHOWN @ U45,
SEE INSTRUCTIONS
BELOW FOR ACTUAL
LOCATION)
EPROM ON NON-FLASH
PG (SHOWN @ U44, SEE
INSTRUCTIONS BELOW
FOR ACTUAL
LOCATION)
U45
For software ver. 1.05 or higher; EPROM's are already installed at both U44 and U45. Follow instructions below.
For PAL ver. 1.00 and 1.01, or if the PAL version number is not marked: Remove the PAL from U50.
Insert the new CHIP SELECT PAL, ver. 1.03, P/N 401228, at U50.
Remove EPROMs from locations U44 and U45
Insert the new BIOS EPROM, P/N 401597, at U44
Insert the PG flash adaptor, P/N 401596, at U45
Remove NVRAM DS1230Y from U49 and NVRAM DS1386 from U54.
Insert NVRAM DS1245Y into U49 and NVRAM DS1248Y into U54.
Install two jumper shunts on J12.
Install two jumper shunts on J13.
Using two lengths of the black 28 AWG wire and referring to Figure 2, connect the PG flash adaptor to the PG main
board. Use silicon glue to hold the wire on the board.
U44
U47
U53
Figure 2
Non-Flash to Flash without EEPROM
U41
U46
U49
DS1245Y
U50
U54
DS1248Y