Thread the brass standoff into the port at the
bottom of the reservoir.
Open the slide valve by sliding the shuttle up:
Valve operation is UP-OPEN and DOWN-CLOSED
when the kit is in the orientation shown in
Figure 1.
Squeeze the hand pump until vacuum is near
25 inHg.
Read the vacuum on the gauge
attached to the hand pump.
Fluke. Keeping your world up and running.®
Fluke Corporation
DH Instruments Division
4765 East Beautiful Lane
Phoenix, AZ 85044-5318 USA
Ph: 602.431.9100
Fax: 602.431.9559
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Figure 3: Evacuating the E-DWT-H with fill kit
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
Close the slide valve by sliding the shuttle
down.
Fill the reservoir cup halfway with the liquid test
media. Refer to E-DWT-H User's Manual for
supported test fluids.
Remove the brass standoff from the reservoir
port. Liquid will be drawn into the evacuated
system.
From other countries: +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access:
http://www.dhinstruments.com
©2009 Fluke Corporation. Specifications subject to change
without notice. E-DWT-H and E-DWT are registered trademarks of
Fluke Corporation, DH Instruments Division.
Printed in U.S.A. 7/2009 3463534A D-EN-N
Modification of this document is not permitted
without written permission from Fluke Corporation.