3A HEALTH CARE ASPEED 2 PROFESSIONAL Manual de instruções - Página 5

Procurar online ou descarregar pdf Manual de instruções para Equipamento médico 3A HEALTH CARE ASPEED 2 PROFESSIONAL. 3A HEALTH CARE ASPEED 2 PROFESSIONAL 9 páginas.

The antibacterial filter must be inserted directly in the pot's lid. Never use the aspirator without the antibacterial filter, since
it is very dangerous for the patient. The collection pot is supplied with overflow valve, lid and pot in a see through
material (polycarbonate). The antibacterial filter must be placed directly in the lid, only on the hole called VACUUM. The
antibacterial filter protects the suction circuit from possible contaminants sucked during use. All the components of the pot can be
conventionally sterilised in autoclave at a maximum temperature of 121°C, or by boiling them for 10 minutes. We recommend
to replace the whole pot every 30 sterilisation cycles. Do not turn over the pot during use to prevent the backflow valve from
being activated; if this happens turn off the aspirator and remove the tube connected to the antibacterial filter. Never use the
aspirator without the secretion collection pot and/or without the antibacterial filter.
Connection: connect one end of the short sterilisable silicon tube to the rubber holder of the antibacterial filter and put
the latter in the "VACUUM" input of the blue lid, connect the other end to the "INLET" input of the aspirator. Connect one
end of the sterilisable silicon tube to the "PATIENT" input of the blue lid; connect the sterile single-use manual regulator to the other
end and connect the single-use sterile pipe to the regulator.
The aspirator can be used with the 1000 cc re-usable transparent secretion
collection pot and with the single-use bag supplied. In this case the antibacterial filter is integrated in the single-use bag, there-
fore the antibacterial filter and the blue lid with the valve should not be used. The filter embedded in the bag, also
prevents the reflux of the liquids sucked towards the aspirator when it is full, or when it is inadvertently turned over. In this case to
restore the device to normal operation, the single-use bag shall be replaced. For the cleaning and disinfecting the vial, sterilise
the single parts in autoclave at a maximum temperature of 121°C, or by boiling them for 10 minutes. The bag is single-use and
it MUST be replaced after each use. The bag must always fully inserted in the pot to avoid vacuum loss.
Note: Maximum use depression of the single-use bag: -0.75 bar.
Connection: connect one end of the short sterilisable silicon tube to the yellow rubber holder (VACUUM) of the lid
and the other end to the "INLET" input of the aspirator. connect one end of the long sterilisable silicon tube to the red rubber
holder (PATIENT) and connect the sterile single-use manual regulator and the single-use sterile pipe to the other end.
• The pipe, the suction flow manual control are sterile single-use products and they must be replaced after each application.
• The antibacterial filter is a sterile single-use product and it must be replaced after each application.
• Never wash the device under the water or immersing it in the water, clean the external cover of the device with a soft cloth
dampened with some (non abrasive) detergent.
ASPEED 2 does not need maintenance and/or lubrication, although it is required to perform some simple checks before each
use:
- Check the integrity of the shell and the power supply cable.
- Using a finger, close the suction connector and verify that the vacuum level reaches 0.80÷0.85 bar.
- Verify that there is no disturbing noise that may be the symptom of a malfunction.
- Check that the cage is correctly positioned inside its housing. It must be aligned with the suction hole of the lid, so the valve
of the float can obstruct it when the sucked liquid is too much against the maximum quantity that can be assimilated by the pot.
- Check that the float is mounted in the right position and that it is free to slide inside the cage (dirt and deposits can obstruct its
movement). The picture below shows how to properly put the float in the cage.