DIALAB DTN-410 K Manual do utilizador - Página 19
Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição DIALAB DTN-410 K. DIALAB DTN-410 K 20 páginas. Photometer
Place a test tube with water in sipper and press pump actuator.
PUMP ACTUATOR
Repeat until absorbance readings are stable within +/- 0.005 Abs.
If readings are not stable, aspirate solution #1 and wait for 2 or 3 minutes. Then aspirate the
same solution again, aspirate soaking solution #2 several times (3 at least). Finally, aspirate
water and verify stability of reading.
Pump Actuator
In Wash method: runs the pump
In Run method: aspirates blank, standard and sample
DIALAB PHOTOMETER DTN-410K USER MANUAL
I 19 I