ABB Endura AZ40 Instruções de funcionamento
Procurar online ou descarregar pdf Instruções de funcionamento para Equipamento industrial ABB Endura AZ40. ABB Endura AZ40 16 páginas. Oxygen and carbon monoxide equivalent (coe) analyzer
Também para ABB Endura AZ40: Manual (16 páginas), Manual de Instruções (20 páginas)
—
A B B M E A S U R E M E N T & A N A LY T I C S | I N S T R U C T I O N | I N S/A N A I N S T/0 0 7- E N R E V. B
Endura AZ40
Sensor assembly
1
Endura AZ40
sensor assembly
This publication details replacement procedures for
the following AZ40 orifice and seals kit (AZ400 752).
Before carrying out any procedures, see Section 3.
The following procedures must be carried out by a
suitably-trained technician:
• Cleaning/Replacing CO and O2 orifices:
Sections 4, 5, 6 and 7 (pages 4, 4, 6 and 7)
• Replacing sensor assembly gaskets:
Sections 4, 5, 6 and 8 (pages 4, 4, 6 and 8)
• Cleaning / Replacing sensor assembly aspirator:
Sections 4, 5, 6 and 8.4 (pages 4, 4, 6 and 12)
• Replacing sensor assembly flow disk:
Sections 4, 5, 6 and 8.5 (pages 4, 4, 6 and 14)
Kit contains:
• Orifices, gaskets, seals
• This publication
Tools required
• Transmitter door key (supplied with transmitter)
• Operating instruction OI/AZ40-EN*
• 3/16 in, 1/4 in, 5/32 in allen keys (wrenches)
• 7/8 in open-ended spanner (wrench)
• 10 in adjustable spanner (wrench)
• Medium flat-bladed screwdriver
• 2 Small screwdrivers
(1 × flat-bladed, 1 × crosshead)
• Single-edged razor blade
• Anti-seize compound
(suitable for temperatures up to 200 °C [392 °F])
Introduction
Orifice and seals kit
Kit reference: AZ400 752
Measurement made easy
*Operating instruction
OI/AZ40-EN
mandatory safety information and can be downloaded
from the link (above) or by scanning this code:
2
For more information
Further information is available from:
www.abb.com/analytical
or by scanning these codes:
Sales
Service
contains