The flow of water to the drain
does not stop.
Check that you have followed ZERO's
operating logic.
ZERO works through an airless dynamic
storage tank to ensure a high instant flow during
dispensing. Therefore, the tank requires a longer
time to refill the 3 liters. The minimum refilling time
will be about 10 minutes and will be affected by the
pressure and temperature of the incoming water
to ZERO. In unfavorable conditions (use at the
extreme end of the minimum conditions for use)
the filling of the tank can take up to 45 minutes. If
you do not know the installation conditions, close
the tap where the purified water comes out and
wait for the drain to stop without operating any
other component of ZERO in any way.
Check the temperature and pressure of
the incoming water.
Check that the temperature of the incoming water
is within the acceptable range (4-38°C) and that
the pressure of the inlet water is a minimum of
2 bar. The operating conditions are also provided
in the installation guide. We recommend calling a
professional technician if you are not able to verify
these conditions yourself.
38°C
4°C
min.
2 bar
ZERO Troubleshooting guide
I cannot change the pre- and
post-treatment filters.
Do not reverse the pre-filter and post-filter.
When changing the filters, it is important not to
reverse their positioning and not to force the
tightening to avoid causing damage to the system.
The two filters have different beaks, so it is
important to insert them in the correct place.
I just installed ZERO and no water
comes out of the drain.
If less than 1 minute has passed, keep waiting.
This might be caused by the presence of air
inside the ZERO circuit, which must be pushed
out through the drain line or purified water line
within the first phase of installation. If less than
one minute has passed after the opening of the
hydraulic supply, wait a few more minutes.
Check that you have removed the cap from the
output drain.
During installation, ensure that you have removed
the cap from the drain and have connected the
pipe correctly. Additionally, if the drain is clogged,
the membranes will not work properly and will
deteriorate very quickly.