Guidance and manufacturers declaration - electromagnetic immunity
The Biomet
SpinalPak
®
electromagnetic environment specified below. The customer or the user of the Biomet
Non-invasive Spine Fusion Stimulator System should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1. At 80 MHz and 800 MHz, the higher frequency applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a F ield strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Biomet
the applicable RF compliance level, the Biomet
System device should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the Biomet
Non-invasive Spine Fusion Stimulator System.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Non-invasive Spine Fusion Stimulator System is intended for use in the
®
IEC 60601
Compliance level
test level
Not Applicable
Not Applicable
3 V/m
80 MHz to 2.5
GHz
SpinalPak
®
®
Table 3
Portable and mobile RF communications equipment
1 V/m
should be used no closer to any part of the Biomet
SpinalPak
including cables, than the recommended separation
distance calculated from the equation applicable to the
where P is the maximum power output rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
Non-invasive Spine Fusion Stimulator System is used exceeds
SpinalPak
®
®
14
Electromagnetic environment -
guidance
Non-invasive Spine Fusion Stimulator System,
®
frequency of the transmitter.
Recommended separation distance
d = 3.5 √ P 80 MHz to 800 MHz
d = 7 √ P 800 MHz to 2.5 GHz
distance in meters (m).
marked with the following symbol:
Non-invasive Spine Fusion Stimulator
SpinalPak
®
®
®
b
SpinalPak
®
®