DiaTecne PulsePen Manual de referência rápida - Página 7

Procurar online ou descarregar pdf Manual de referência rápida para Monitor de tensão arterial DiaTecne PulsePen. DiaTecne PulsePen 7 páginas.

Traffic Lights (software)
The small traffic lights graph, at the top of the computer screen during the signal capture, is updated each
cardiac cycle. If it doesn't occur, it means:
The R wave of the ECG is not detected (in the case of Ecg + Tonometry).
The foot of Tonometer - sensor1 curve is not detected (in the case of 2 x Tonometry, ETT model only).
Possible causes
The ECG amplitude is too low
The R wave is reversed (downwards)
The R wave has pathological shape
Tonometer1 (red curve) is off and/or no signal is
applied
Captured signal's amplitude on Tonometer1 is too low
DiaTecne
PulsePen -

Quick Reference Guide

Ecg + Tonometry
Re-position one or both the ECG electrodes in order to
obtain signals with higher amplitude.
Exchange the Ecg cables or electrodes.
Proceed with the capture for at least 10 seconds, save
the signals and manually add markers on the R waves.
2 x Tonometry
Switch on the Tonometer1 and position it on the artery
under investigation (Carotid).
Re-position the Tonometer1 in order to obtain signals
with higher amplitude.
7
1.0
Remedy