3.3.3 Unit
This sub-menu serves to select the parameter unit: for metric (µm) or inch (µin) indication /
measurement.
3.3.4 Nx
Switching-on/off of the roughness classes N1 to N12
3.3.5 Amplification
This sub-menu allows the selection of the amplification of the instrument. Normally, the setting 'auto'
will be sufficient. The instrument itself will then select, whether the amplification for small or large
roughness values is to be set. For very irregular surface or for surfaces with deep grooves, the
amplification can be set at choice with factor 'x1' or 'x10'.
3.3.6 Switch-off
In this sub-menu the duration of time can be selected, within which the instrument should be switched-
off (in minutes, in steps of 30 seconds).
3.3.7 Factory configuration
When selecting this menu point, the actual instrument setting will be reset to the original factory
setting. The view of menu (page 11) shows the basic settings (Standard).
3.4
Memory functions and printout of memorized values
Select menu 'memory functions'. If the memory function is active, an 'M' appears in the display.
15 measuring profiles can be memorized. See 2.4, page 8, Memory function
With 'on'/'off' the function is switched-on or –off. After each measurement effected the measuring
value can be saved.
With 'clear' the complete memory contents are cleared.
With "USB" the stored values are released to the Interface USB. The data may be taken over with a
suitable software. It's necessary to provide the PC with an USB driver.
3.5
Setting of tolerance
The roughness unit (Ra, Rt, Rz) for which a tolerance should be set, may be selected here. Set the
maximum tolerance value. If this value is exceeded, the indication of this value will blink on the
display. This sub-menu serves also to set the c-value for the calculation of the RPc unit.
3.6
Rechargeable battery
Here, the charge of the accumulator in indicated. This information appears in the screen also each
time after switching-on the instrument. In order to secure a long lasting battery-powered operation, the
instrument is provided with an automatic switch-off device. If no key has been actuated within a
selected duration of time, the instrument switches-off. This duration of time of 1 ... 10 minutes until
switch-off may be set in the menu 'Config'.
3.7
Date and clock- time
The DIAVITE COMPACT is equipped with an integrated clock (time and date).
The clock is integrated in the battery pack. If this has to be replaced or if it had been separated from
the instrument, the time and date will have to be reset.
3.7.1 Indication of date and clock-time
In the first sub-menu, the actual time and the date can be indicated.
3.8.2
Setting of time and date
This menu option allows the setting of the time. With key 'Lc', the cursor can be moved to right, with
key 'P' to left.
The keys 'R' and 'Lt' allow setting the time values under the cursor, i.e. hour, minute and date.
With 'Esc', the values set are taken over. Press button 'M' to return to 'Menu'.
3.9
Indication of the version
In this option, the versions of the hardware and of the software incorporated is indicated.
DIAVITE COMPACT
11