2
density units (µW/m
, mW/m
or µW/cm
) using
the standard far-field formulae for electromagnetic
radiation.
The conversion is invalid for near-field
measurements, as there is no generally valid
relationship between electrical and magnetic field
strength in this situation. Always use the default
units of the sensor when making near-field
measurements.
8.4 Result modes
The bar graph display always shows the
instantaneous measured dynamic range value.
The digital display shows the result according to
one of three modes, which can be selected.
Instantaneous : The display shows the last value
measured by the sensor, no symbol is displayed.
Instantaneous mode is the default setting when
the meter is turned on.
Maximum instantaneous (MAX) : The digital
display shows the highest instantaneous value
measured so far, the "MAX "symbol is displayed.
Average (AVG) : The digital display shows the
12