Operating Instructions, Ctd.
7. Automatic Shutoff This instrument will
automatically shut down after "" 34 minutes if
no keys are
pressed.
8. Automatic Shutoff Override
You may override the automatic shutoff
fea-
ture at startup by pressing down the "Hold"
key. and the "On" key at the same
time.
The
letter 'T' will light up in the
display,
indicating
that the shutoff timer is in the override
mode.
T
9. Minimum/Maximum
Press "Min/Max" once and the minimum val-
ues will
be
displayed for approximately 7 sec
onds.
Press "Min/Max" again during this time and
the r:naximum values will be displayed for 7
seconds.
Press "Min/Max" a third time during this peri-
od and the unit will return to normal opera
tion.
Unit will always return to normal operation
after 7 seconds if no keys are pressed.
Note: Turning the unit off resets the minimum
and maximum values
.
a
Minimum Display
Maximum Display
Battery Installation
Battery should be replaced as soon as possible
after low battery indication to maintain rated
accuracy of unit
Calibration
Your instrument was carefully tested
and calibrated before being
shipped from the
factory.
Additional calibration is not
required. However. should calibra-
tion be desired in the future. follow
these
procedures:
1 . Identify a standard that is more
accurate than this unit Factory
supplied calibration salts or
elec-
tronic instruments recently
cali-
brated at a certified lab are rec-
ommended. Sling psychron-
meters and instruments using
mechanical sensing elements
(human hair. wood
elements.
etc.) are NOT
recommended. These instru-
ments can be subject to great
inaccuracies due to the nature
of
mechanical
measurement.
2. Place the probe of the unit into
the calibration salt capsule or
into a controlled environmental
chamber and allow for complete
stabilization. For best
results.
calibrate the unit at levels typical-
ly monitored during normal
a
operation. Calibration in an
open room without the salt cap-
sule or a chamber is not recom-
mended as temperature and
humidity can vary greatly within
a very small
area.
3. To activate the calibration mode.
turn the unit
off.
Now press the
"On/Off' key and the "C/F" key at
the same time. A small
"J"
will
appear in the display to indicate
you are in the calibration mode.
Step 3: Activating Calibration
Mode