Alpha Moisture Systems dewSMART ADSH Manual de instruções - Página 4

Procurar online ou descarregar pdf Manual de instruções para Instrumentos de medição Alpha Moisture Systems dewSMART ADSH. Alpha Moisture Systems dewSMART ADSH 15 páginas. Dewpoint meter

J-
~
~
Page 3
INST ALLA TION
INSTALLING THE INSTRUMENT INTO A PANEL
Make a cut-out in the donor panel 92.0/92.8 x 45.0/45.6mm (DIN 43700).
The maximum panel thickness is 10mm and, if an effective IP65 weatherproof seal is
required, the minimum recommended panel thickness is 1.6mm.
Pass the instrument cabinet through the cut-out in the donor panel and slide the panel
clamp over the instrument, from the back, ensuring that the narrow section is at the top.
Screw the panel clamp retaining screws into the instrument cabinet until the instrument is
clamped in position. The screws must be tightened sufficiently to effect a seal between
the front of the donor panel and the back of the instrument bezel, but never over tightened
to the point of fracturing the panel clamp or instrument case.
FREE STANDING CABINET (If Applicable)
Remove the 4 screws at the sides ofthe cabinet and remove the cover.
Remove the plug-in connectors from the back of the instrument.
Pass the instrument
through the cut-out in the front of the cabinet chassis and slide the panel clamp, from
inside the chassis, over the back of the instrument, ensuring that the narrow section is at
the top.
~
Ensure that the connector sockets at the back of the instrument are correctly located in the
cut-outs provided.
Enter a No.2 pozidriver through the 2 round holes at the back of the chassis and screw
the panel clamp retaining screws until the instrument is firmly clamped in position in the
chassis.
Refit the cabinet cover, replace the retaining screws and refit the plug-in connectors.
The plug in cable connectors are NOT interchangeable, therefore ensure
that the three 5 pin connectors are kept in the appropriate slots.
NOTE
INSTRUMENT WIRING
~
C
Please refer to the electrical connections diagram on page 11.
POWER SUPPLY
Connect the power supply cable to the orange coloured terminal block and observe
polarity.
pi-..
The power supply should be between 90 and 250V AC @ 50/60Hz.
~
-,
NOTE
Wires are retained by screws. Ensure that the exposed section of the wire
is fully inser1ed and that no loose strands are exposed.
.......
-
J
DOC:
DsrOOOIDT20.3-J-O