Axium Clinical Programmer Manual do utilizador - Página 6
Procurar online ou descarregar pdf Manual do utilizador para Equipamento médico Axium Clinical Programmer. Axium Clinical Programmer 16 páginas.
Change the Date and Time
To change the date, select the drop down arrow on the
right side of the "Set Date" box. A calendar will appear
and you can set the month, day and year using your
stylus.
To change the time (24 hour format), first select the
hour or minute field that you would like to change.
To change the selected field, use the "Up" or "Down"
arrows to increase, decrease or toggle the setting.
NOTE: Establishing a connection updates the NS
device's clock to the newly set time.
Switch to FCE Programmer
By checking this box the Clinical Programmer will get
additional functionality, which should only be used by
Spinal Modulation's Field Clinical Engineers and Staff.
Set Password
A password may be set to limit access to the
Programmer.
The password is for the Programmer itself and is not
associated with any NS.
Establishing Communication with the NS Device
To change the patient's stimulation settings, you must
first establish communication between the Clinical
Programmer and the patient's NS device.
1. Make sure that the Clinical Programmer is turned on
and the Main Menu screen is displayed
2. Press "Connect to Stimulator" on the Main Menu.
3. Select the text box next to "Stimulator SN:"
4. Enter the serial number using the pop-up keyboard.
If the serial number format is valid for a NS device, the
"Connect" button will be enabled.
5. Press the "Connect" button.
After pressing the "Connect" button, the "Cancel" button
becomes enabled. If the Cancel button is pressed, the
telemetry connection is cancelled.
7
6. Move the Clinical Programmer magnet over the NS device in a circular motion to connect. The
indicator status bar on the bottom left of the screen will display "Connected" if the connection
attempt is successful. If the Programmer could not communicate with the NS device, an error
message will appear and "Disconnected" will be displayed in the status bar.
NOTE: If after 2 minutes the Clinical Programmer has failed to communicate with the NS device,
the programmer will automatically cancel the connection attempt. Try to communicate with the
NS device by pressing the "Connect" button again moving the magnet symbol on the Clinical
Programmer over the NS device in a circular fashion.
When a successful connection is established, the Programmer chimes and the NS device will be queried.
7. For the duration of the programming session keep the Clinical Programmer within 3 feet of the NS
device. Moving the programmer too far away may cause telemetry connection to be lost.
Back to Main Menu
Located at the bottom right side of the Programmer Connect window, the "Exit" button is used to return to
the Main Menu.
Navigation and Screen elements
Neurostimulator Dashboard
Once the selected NS device is connected to
the Clinical Programmer, the NS Dashboard is
displayed in the screen's header providing:
• Patient ID: The patient's ID Number
• Stimulator Serial Number: The NS device's
serial number
• Alerts Button: The button turns orange when
any of the NS System Alerts become active.
When the "Alerts" button is orange, press the
button to display a window showing details
of all the System Alerts. An example of the
screenshot is shown to the right.
8