Use  a  2.5mm  Allen  wrench  or  driver  to  remove  the 
three  M3x12 hex, socket‐head  cap screws securing the 
guide  block  to  the  back  side  of  the  focus  block  (see 
Figure 13).   
Figure 13 – Remove screws securing the guide block 
Separate  the  (now  loose)  guide  block  from  the  focus 
block (see Figure 14).   
Figure 14 – Remove the guide block from the focus block 
The components of the focus block assembly are shown 
in Figure 15. 
Figure 15 – The components of the focus block assembly 
Remove the two brass shims from the guide block (see 
Figure 16). 
Complete Teardown, Cleaning, and Reassembly of the Olympus BH‐2 Sliding Focus Block  
 
Figure 16 – Remove the brass shims from the focus block 
Use a 2.5mm Allen wrench or driver to unscrew the two 
M3x16  hex,  socket‐head  cap  screws  from  the  wedge 
nuts,  and  remove  the  (now  loose)  wedge  nuts  (see 
Figure 17). 
 
Figure 17 – Remove screws and wedge nuts 
Clean the Various Slide Components 
 
Disassemble  and  clean  the  various  components  of  the 
sliding  focus  block  mechanism  per  the  following 
procedure. 
Remove the Guide Rails from the Arm Casting 
Remove  the  four  long  guide  rails  from  the  two  side 
grooves in the arm casting (see Figure 18).  
 
Figure 18 – Remove four guide rails from the arm casting 
Revision 2 