8-2 Detailed Assembly Procedures
@
²
Fig. 32
³
Fig. 33
ƒ
|
Fig. 34
1
Attaching the Halogen Bulb
}The applicable lamp bulb is the 12V100WHAL-L (PHILIPS 7724) or the
12V50WHAL-L (LIFE JC).
1. Fully loosen the clamping screw @ at the top of the lamp housing using
the provided Allen screwdriver.
2. Remove the lamp housing ² by lifting it up.
3. Tilt the bulb socket ³ by 90° in the direction of the arrow.
4. While pushing down the bulb clamping levers | down, hold the halo-
gen bulb ƒ with gloves or a piece of gauze, insert the bulb pins ... into
the sections † as far as they will go.
Then return the lamp clamping lever gently back to the original position
to clamp the bulb.
!To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
5. Fit the lamp housing from up and tighten the clamping screw @ by
applying downward pressure.
!Caution for Bulb Replacement During or Right After Use
The bulb, lamp housing and areas near these will be extremely hot
...
during and right after use.
Set the main switch to "
the wall outlet, then allow the old bulb and lamp housing to cool
before replacing the bulb with a new one of the designated type.
(Fig. 32 - 34)
" (OFF), disconnect the power cord from
BX51M