Olympus CH20 Manual de reparação - Página 7
Procurar online ou descarregar pdf Manual de reparação para Microscópio Olympus CH20. Olympus CH20 31 páginas. Biological microscope
CH10/CH20
Items
4
Focusing
mechanism
5
Condenser
holder
6
Condenser
7
Collector lens
part
8
Light source
9
Filter
10
Weight
11
Storing case
A. OUTLINE OF PRODUCT
Specifications
4) Working range • 40 mm in longitudinal
‚ 76 mm in lateral direction
5) Clip holder (lever type)
ž Holding capacity: a standard slide glass (76 ×
26 mm)
ž Specimen thickness: 0.9 - 1.4 mm
ž Inclination form: 30°
6) With biological clip holder (except FM type)
1) Coaxial coarse and fine adjustment knobs
2) Coarse / fine adjustment stroke: 20 mm or
more
3) With fine scale: 2.5 µm / graduation
4) Fine adjustment stroke per rotation: 0.3 mm
5) Coarse adjustment knob (with the torque
adjustment mechanism)
6) Without coarse adjustment lock mechanism
1) Dovetail type (Condenser can be attached and
removed freely.)
2) Rack and pinion vertical motion, rotating; knob
type, stroke: 10 mm or more
1) Numerical aperture: N. A. 1.25 (using in
immersion oil)
2) Aperture diaphragm: diaphragm diameter; φ1.7
- 29 mm
3) Exclusive condenser with dovetail and rack
integrated
4) A filter of φ32.5 can be attached (CH2-FH).
5) CH2-RS10, RS40 and DS can be attached.
6) CH3-AL can be attached (CH20F only).
1) CH3-FS can be attached: circular fitting snap
type
2) A filter of φ45 can be attached.
1) 6V20W halogen lamp
2) Lamp replacement by opening and closing back
cover
3) Built-in transformer, phase control system
4) Dial continuous light intensity control:
indicating the position by graduation
Blue filter (KB-P) is attached directly to the lower
part of the condenser.
CH20F: 4.4 kg
CH10F: 3.4 kg
ž Wooden box (CH20)
ž Plastic case (CH10)
direction
A-3
Remarks
Six diaphragm
blades
Optional
Optional
Optional
Halogen bulb:
Philips 7388
(average life-
time: 100 H)
at CH10/20BIMF
setting