Olympus CK40M FRAME Manual de reparação - Página 6

Procurar online ou descarregar pdf Manual de reparação para Microscópio Olympus CK40M FRAME. Olympus CK40M FRAME 38 páginas. Compact inverted metallurgical microscope

Olympus CK40M FRAME Manual de reparação
CK40M
(4)
This model can be combined with the following observation tubes:
CH3-BI, CH3-TR45, CK40-TBI
4.

Specifications

Item
1
Focus
2
Revolving
nosepiece
3
Stage
Illumination
system
5
Power supply
6
Weight
A. OUTLINE OF PRODUCT
• Revolving nosepiece vertical movement (Stage fixed type)
• Coaxial coarse/fine adjustment knobs (Coarse adjustment with
torque adjustment mechanism)
• Roller guide type
• Stroke: 9 mm from the focus position at 1 mm above stage top
7 mm upward and 2 mm downward
• Stroke by 1 rotation of coarse adjustment knob: 36.8 mm
• Stroke by 1 rotation of fine adjustment knob: 0.2 mm
Quadruple
Plane stage
160 (L) × 250 (W) mm, stage insert plate provided (no accessory)
A mechanical stage and stage clips can be mounted.
Light source
6 V 30 W halogen bulb (6 V 30 W HAL)
Average service life of bulb: 100 hours
U-LS30-3-4 lamp socket
• Aperture diaphragm (AS)
Lever control type
Adjustable from φ15 mm (minimum) to φ55 mm (maximum)
• Filter slot
Drop-in type
φ25 mm filter: Up to thickness of 25 mm
• Polarizer slot
Drop-in type
• AS projection magnification (AS → Assumed objective exit
pupil): 1.71X
• Light source projection magnification (Filament → Assumed
exit pupil): 3.11X
Continuous light intensity control by variable resistance
Rated output voltage: 6 V
Built-in rated input voltage selector switch
100/110-120 V for F100
220/230-240 V for F200
Rated input frequency:
50 Hz, 60 Hz
Built-in fuse
Fuse rating: 0.6/0.3 A
Rated load:
6 V 30 W halogen bulb
(PHILIPS 5761)
Power consumption: 72 VA (W)
7.5 kg
Specifications
A-3
Remarks
Maker: Philips
Model No. 5761
Cover provided
(Filter/polarizer