Olympus CX31-P Manual de instruções - Página 17
Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus CX31-P. Olympus CX31-P 36 páginas. Polarizing microscope
4-6 Condenser
3
1
Fig. 17
Field iris diaphragm image
Field of view of eyepiece
Fig. 18
2
Fig. 19
1
Centering the Field Iris Diaphragm
}When using the U-CTAD centering adapter for objective lens, perform the
operation in
section 5-1 (page 17) before proceeding to the following steps.
1. Engage the 10X objective in the light path, bring the specimen into focus
3
and rotate the field iris diaphragm ring 1 counterclockwise to stop down
the field iris diaphragm slightly.
2. Turn the condenser height adjustment knob 2 so that the image of the
2
field iris diaphragm is focused on the specimen surface.
3. Turn the two condenser centering knobs 3 alternately to make the im-
age of the diaphragm concentric with the field.
4. Open the field iris diaphragm so that its image inscribes the field of view.
If the image is eccentric, adjust the centering knobs again.
5. Enlarge the iris diaphragm image until it just circumscribes the field of
view.
2
Using the Aperture Iris Diaphragm
}The aperture iris diaphragm adjusts the numerical aperture of the illumi-
nation system. Aligning the numerical aperture 1 of the illumination sys-
tem with that of the objective in use improves the contrast and resolution
as well as the focal depth of the observed image.
1
How to adjust the aperture iris diaphragm
Set the aperture iris diaphragm lever 2 of the condenser so that its
numerical aperture 1 matches the NA data indicated on the objective.
As the aperture iris diaphragm lever has a certain width, set it so that the
numerical aperture matches at the center of this width.
When the 100X oil-immersion objective is used, set the aperture iris
diaphragm lever 2 to a numerical aperture of 0.9.
}As microscopic specimens in general have low contrast, it may be opti-
mum to set the aperture iris diaphragm to between 70% and 80% of the
aperture number of the objective.
# Stopping down the aperture iris diaphragm excessively may result
in production of ghost.
3
Adjusting the Centering Adapter for Objective Lens in
CX31-P
(Figs. 17 & 18)
(Fig. 19)
13