4
Assembly
1
Attaching the B excitation fluorescence illuminator
10
b
e
c
a
d
Be sure to set the main switch
CAUTION
fluorescence illuminator to
removing the B excitation fluorescence illuminator.
1
Remove the cover plate for insertion portion of the fluorescence
illuminator by inserting the flathead screwdriver into the notch
Be sure to store the cover plate in a safe place.
NOTE
2
Insert the B excitation fluorescence illuminator into the insertion portion
of the fluorescence illuminator with the brightness adjustment knob
c
facing up, and push it completely until it touches the end.
Removing the B excitation fluorescence illuminator
1
Insert the allen wrench provided with the microscope frame into
d
the hole
at the tube attaching portion of the microscope frame,
and push the stopper
illuminator to pull out the B excitation fluorescence illuminator.
When removing the B excitation fluorescence illuminator,
NOTE
attach the cover plate to prevent dust.
a
of the B excitation
(OFF) before attaching or
e
attached to the B excitation fluorescence
b
.