Using the High/Low Zoom Magnification
3
Stopper
The magnification can be limited in the desired range by
TIP
setting each stopper on the microscope body (also used as
the magnification indices) and the stopper ring on each zoom
adjustment knob. The right zoom adjustment knob is used to
limit the higher end of the magnification, and the left knob is
used to limit the lower end.
1
Using an Allen wrench, loosen the clamping screw of the right stopper
ring a and free it. (Also free the left knob in this way.)
2
Rotate the right zoom adjustment knob b to align the desired higher-
limit magnification with the index (stopper) c .
3
Gently apply the stopper ring a to the stopper c and tighten the
clamping screw using the Allen wrench to retain this condition.
4
Set the left zoom adjustment knob according to the desired lower-limit
magnification by rotating the left knob and clamping the left stopper
ring in the same way as above.
4
Using the Eye Shades
The eye shades are not provided with ESD protection
NOTE
performance.
When ESD protection is required, do not use the eye
shades.
When Wearing Eyeglasses
Use with the eye shades in the normal, folded-down position. This will
prevent the eyeglasses from being scratched.
When Not Wearing Eyeglasses
Extend the folded eye shades in the direction of the arrow to prevent
extraneous light from entering between the eyepieces and eyes.