Olympus SZX-AS Manual de instruções - Página 26
Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus SZX-AS. Olympus SZX-AS 34 páginas. Research stereomicroscope system
9-4 Revolving Nosepiece SZX-2RE
1
External View
Nosepiece mount
2
Assembly
Fig. 29
ƒ
|
Fig. 30
1. Attach the hand guard ² to the revoling noseplece @ by tightening the
³
provided 2 clamping screws ³ (M3 –– 4 mm long) using the provided
Allen wrench (for M3 screws). (Fig. 29)
2. Remove the observation tube and intermediate attachment from the zoom
²
microscope body.
3. Remove the zoom microscope body and focusing assembly from the
pillar, and place them upside down on a flat desk surface.
@
# Place a soft sheet made of rubber or similar material on the desk
surface.
4. Remove the objective from the zoom microscope body and remove the
objective mount ƒ by loosening the 3 clamping screws | using the
provided Allen wrench (for M3 screws). (Fig. 30)
Hand guard
Objective mounts
Clamping
screws
SZX
Allen wrench
(for M3 screws)
M3 –– 4 mm long (x 2)
M3 –– 6 mm long (x 3)
(Figs. 29 - 32)
22