Olympus SZX-R Manual de instruções - Página 3
Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus SZX-R. Olympus SZX-R 34 páginas. Research stereomicroscope system
2
Maintenance and Storage
1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze slightly
moistened with a mixture of ether (70%) and alcohol (30%).
Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to keep
these chemicals away from open flames or potential sources of electrical sparks -- for example, electrical equip-
ment that is being switched on or off. Also remember to always use these chemicals only in a well-ventilated
room.
2. The equipment uses plastic resins extensively in its external finish. Do not attempt to use organic solvents to clean the
non-optical components of the microscope. To clean these components, use a lint-free, soft cloth lightly moistened with a
diluted neutral detergent.
3. Never disassemble any part of the microscope as this could result in malfunctions or reduced performance.
4. When not using the microscope, keep it covered with the dust cover provided.
Caution
3
If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition,
the microscope may also be damaged. Always use the microscope as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).
4. Only one intermediate attachment can be used. Using two attachments
may obscure a part of the image. However, the vertical illuminator (SZX-
ILLC) is not considered to be an attachment.
}When more than one intermediate attachment is to be combined, they
should be stacked according to the following order, from the bottom to
the top: SZX-ILLC, SZX-AS, SZX-RFA (SZX-RFL), SZX-SDO, SZX-APT, SZX-
BS, SZX-DA, SZX-FAD, then SZX-EPA .
5. The desk surface inclination with respect to the horizontal surface should
be less than 5°.
ii