Mounting the Eyepiece Micrometer Disk
A variety of eyepiece micrometer disks a can be inserted
TIP
into the WHSZ10X-H, WHSZ15X-H and WHSZ20X-H
eyepieces. Please purchase an eyepiece micrometer disk
with a diameter of 24 mm and thickness of 1.5 mm.
1
Unscrew the micrometer disk frame b by turning it counterclockwise
to remove it from the bottom of the eyepiece.
2
Clean the eyepiece micrometer disk a to remove dust and dirt, then
place the disk into the micrometer disk frame b so that the side with
reticule c faces downward.
3
Attach the micrometer disk frame b containing the eyepiece
micrometer disk a by gently screwing it onto the eyepiece d .
· The micrometer disk frame may be tight for certain micrometer
TIP
disks. In this case, turn the frame by holding the circumference
with a light, uniform force or by applying the frame against a
rubber sheet. Do not grasp the frame with a strong force, as this
may deform the frame and make it harder to remove it.
· Be careful not to touch the lens surface with your finger.
The WHSZ20X-H eyepiece is designed to apply a
NOTE
magnification to the focusing plane of the micrometer
disk.
As the magnification coefficient is 1.35, be sure to perform
magnification compensation when using this eyepiece in
measurement.
When the micrometer disk is engaged in their light path,
the light path length will extend and the diopter scale may
deviate toward the + direction from the normal indication.
However, this does not pose any problem in actual
observation.
When the eyepiece micrometer disk a is not used, wrap it in
TIP
a clean sheet of soft paper before storage.