Olympus SZX10 Manual de instruções - Página 33

Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus SZX10. Olympus SZX10 40 páginas. Research high-class stereo microscope
Também para Olympus SZX10: Visão geral (24 páginas), Manual de instruções (36 páginas)

Olympus SZX10 Manual de instruções

9-5 BX Stage Adapter Type 2 SZX-STAD2

TIP
1
Assembly
BX stage adapter type 2
SZX-STAD2
NOTE
Downloaded from
ManualsNet.com
This adapter is for installing the U-SIC4R2/SIC4L2 large stage* on the SZX2-STL large stand**. When this
stage adapter is used, the SZH-P400 auxiliary pillar should be used to cover the height of the stage adapter
(about 125 mm).
In addition, when a low-magnification objective with a long working distance (DFPL0.5X-4) is used, the
SZH-P600 auxiliary pillar should be used (always together with the drop prevention collar).
The U-SVL or S-SVR BS stage for the BX series microscopes can also be mounted but the operabilitydeteriorates.
*
The U-SVLB and U-SVRB cannot be used due to their long stage knobs.
** STAD mount for ILLT SZX2-STADM is necessary for using the SZX2-ILLTQ/ILLTS LED transmitted light illumination
base. In addition, the LED transmitted illumination built in SZX2-ILLTQ/ILLTS cannot be used for observation.
For details, refer to the instruction manual for the SZX2-ILLTQ/ILLTS LED.
Frosted filter
2
Mirror knob
Large base
SZX2-STL
For simplified transmitted light observation, place the mirror knob on the front and use a frosted filter.
search engine
Large stage
U-SIC4R2/SIC4L2
3
Allen wrench
(provided with the base)
Clamping screw
1
Mounting threaded hole
Mounting threaded hole
SZX10
29