SPECIFICATIONS

Item
1. Zoom microscope body
· SZX-ZB12
· SZX-ZB9
* Total magnification with a 1X objective
and 10X eyepiece.
2. Focusing assembly
· SZX-FOF
· SZX-FO
3. Observation tube
· SZX-TR30
· SZX-BI30
· SZX-TBI
4. Standard base
· SZX-ST
5. Large base
· SZX-STL
15
SZX-ZB12
Left/right zoom magnification system.
Zoom drive system: Horizontal knob.
Click stop ON-OFF switchable per zoom magnification.
Zoom ratio: 12.8 (0.7X to 9X)
Magnification indications*:
7, 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 90
Objective mount: Threaded mount
Built-in aperture iris diaphragm.
SZX-FOF
Focusing system: Rack & pinion roller guide
(with coarse focus adjustment knob tension adjustment ring).
Built-in counter balance.
Coaxial coarse/fine focus adjustment
knobs.
Coarse focus adjustment knob stroke: 80 mm
Coarse knob stroke per turn: 36.8 mm
Fine focus adjustment knob stroke: 80 mm
Fine knob stroke per turn: 1.5 mm
SZX-TR30
Trinocular
Tube inclination: 30°
Light path selection: 2 steps
(Bi 100%, Bi 20%:Photo
80%)
Interpupillary distance adjustment: 50 to 76 mm.
Eyepiece clamping knob provided.
Eyepiece: WHS series eyepiece.
Pillar height: 270 mm.
Base dimensions: 300(W) x 260(D) x 30(H) mm.
Specimen holder attachable.
With stage adapter mounting holes.
Pillar height: 400 mm.
Base dimensions: 400(W) x 350(D) x 28(H) mm.
Specimen holder attachable.
With stage adapter mounting holes.
With drop prevention collar (SZX-R).
Specification
SZX-ZB9
Zoom ratio: 9 (0.63X to 5.7X)
Magnification indications*:
6.3, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 57
Use of AS unit (SZX-AS).
SZX-FO
–––
Coarse focus adjustment knob only.
Coarse knob stroke per turn: 21.2 mm
–––
SZX-BI30
Binocular
Tube inclination: 5° to 45°
–––
SZX-TBI
Tilting binocular