7-3 Electrosurgery
The type of output current and its intensity must be determined by refer-
ence to related papers and based on one's own experience.
1
Turn ON the power switch.
2
Press the desired current selector switch (CUT, BLEND or COAG).
3
Power Feeding
Selection of Current
If none of the current selector switches are pressed, none of the current
indicators above them will be lit; instead, the warning tone sounds and
current will not be delivered when the footswitch is depressed.
If more than two current selector switches are pressed, the current indicators
corresponding to the pressed switches will light: in this case also, the warn-
ing tone sounds and the output is inhibited.
Selection of Output Level
Press the desired output adjustment switch.
If none of the output adjustment switches are pressed, none of the LEVEL
indicators above them will be lit; instead, the warning tone sounds and
current will not be delivered when the footswitch is depressed.
If more than two output adjustment switches are pressed, the LEVEL
indicators above them corresponding to the pressed switches will light;
in this case also, the warning tone sounds and the output is inhibited.
Non-Controlled Document