Olympus UHI-2 Manual de manutenção - Página 19

Procurar online ou descarregar pdf Manual de manutenção para Equipamento médico Olympus UHI-2. Olympus UHI-2 38 páginas.

Olympus UHI-2 Manual de manutenção

1-6 Noise is generated in the unit. (Noise check)

Noise check
Is noise generated during
halt of operation?
Is noise generated during
gas feeding?
End
Gas leaks in the high pressure part.
Check the joints of each pipe.
Primary pressure
reducer safety valve?
Electric-
pneumatic ratio
valve?
Manifold U?
Note 1: With the characteristic of the electric-pneumatic ratio valve, if gas is fed
with the connector opened, the safety valve may function. This is not a
defect.
2-7
TROUBLESHOOTING
Looseness?
Hole?
Replace the primary
pressure reducer.
Note 1
Replace the electric-
pneumatic ratio valve.
Replace the manifold U.