OLYMPUS CIX100 Cleanliness Inspector System
3
Setup of OLYMPUS CIX100 Cleanliness Inspector System at
time of delivery
The CIX100 system is delivered in a state ready for the inspection of dark and
reflective particles on white, matt filter membranes.
The standard filter holder for membranes of 47 mm diameter is included and can
be used directly with a filter membrane. Also, all other filter holders where the filter
is not covered can be directly used in your CIX100 system without recalibration. The
sample type is referred to as uncovered membrane. Additional filter holders can be
ordered. Two filter holders can be mounted to your CIX100 system simultaneously.
For convenient use of samples with different reflectivity than a matt white filter
membrane, e.g. made of cellulose nitrate, it is sensible to redo the white balance
calibration. Whether a white balance recalibration is recommended can be seen
when the filter membrane is inspected in the regular inspection workflow. If the
surface of the filter membrane has a bluish cast, the white balance calibration
should be redone on a clean section of the filter membrane.
No other change in the software settings is to be done for the inspection of
uncovered membranes.