Olympus CKX41 Manual de reparação - Página 29
Procurar online ou descarregar pdf Manual de reparação para Equipamento de laboratório Olympus CKX41. Olympus CKX41 43 páginas. Reflected fluorescence system
Também para Olympus CKX41: Visão geral (7 páginas), Brochura e especificações (4 páginas), Manual de instruções (36 páginas)
CKX31/CKX41
(3) Optical axis (CKX41)
<Fig.2>
1) Remove the cover-L and cover-R.
Screw: 3PUTB2X4SB 3pcs. each (*1)
2) Remove the cover.
Screw: CUTB3X10SA 2pcs. (*2)
Adjustment (B)
Adjustment (A)
Adjustment (A) : Loosen the screws(*3) and adjust the revolving axis by moving the right sleeve.
Adjustment (B ) : Loosen the screws(*4) and adjust the left /right optical axis by moving the left
Adjustment ( C) : Loosen the screws(*5) and adjust the absolute optical axis in the right sleeve
(4) Parfocality <CKX41>
*In the same manner as CKX31, adjust the paforcality by inserting washer under the lens frame-1.
(For adjustment procedure, refer to D-12.)
Insert the jigs in the right sleeve and check
the parfocality.
Jigs:
1) Focusing telescope (FT-36 or U-FT)
2) Standard eyepiece (KN0048; Adapter-1)
3) Standard objective (KN0041)
D. REPAIR PROCEDURE
*4
*5
*3
Screw: CUK3X6SB 2pcs
sleeve. (the right sleeve is taken as a standard)
Screw: CUK3X6SB 2pcs. (*4)
by moving the position of binocular unit.
Screw: CUKK3X6SA 4pcs. (*5)
Cover-L
*1
Cover-R
<Fig.2>
Adjustment (C)
(*3)
.
<Left sleeve>
*2
(Positions of
screw holes)
After checking it in the right sleeve, turn the
helioid to focus on the specimen in KN0041,
then loosen the screws(*1) ,only move the
scale ring and fix it at the position where
scale "0" is aligned to the index.
Note: When using the screw(*2) for fixing eyepiece;
mount the diopter ring with (A) and (B) screw
hole positions aligned without the above
adjustment.
D-2
Cover
*2
Diopter ring (A)
*1
Helicoid ass'y (B)