Olympus CKX41 Manual de reparação - Página 40
Procurar online ou descarregar pdf Manual de reparação para Equipamento de laboratório Olympus CKX41. Olympus CKX41 43 páginas. Reflected fluorescence system
Também para Olympus CKX41: Visão geral (7 páginas), Brochura e especificações (4 páginas), Manual de instruções (36 páginas)
CKX31/CKX41
Parfocality adjustment amount (CKX41: reference value)
Reading value
(before adjustment)
AB129900 AB130000 AB130200 AE092100 AE092200 AE092300
2.0
1.9
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
-0.7
-0.8
-0.9
-1.0
-1.1
-1.2
-1.3
-1.4
-1.5
-1.6
-1.7
-1.8
-1.9
-2.0
Reading value (before adjustment): Reading value of parfocality difference on the helicoid scale of standard eyepiece
Reading value (after adjustment): Theoretical reading value of parfocality difference after adjustment
Ex.) Parfocality differnce is 2.0 mm: when the total thickness (1.05mm) of washers, (0.05mm: 1pc.) and ( 1mm: 1 pc.),
* Numeric character (1 or 2) in the column of washer selection shows the number of washers.
D. REPAIR PROCEDURE
Washer selection
0.05
0.1
0.3
1
1
1
2
1
2
2
2
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
are inserted , parfocality difference becomes -0.03mm reading value.
< standard : 0.7 >
0.7
0.5
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
D-13
Unit: mm
Total
Reading value
washer
(after adjustment)
thickness
1.0
1.05
-0.03
1
1.05
1
-0.13
1
1.00
0.06
1
1.00
-0.04
-0.14
1
1.00
0.95
0.05
-0.05
0.95
0.90
0.14
0.90
0.04
0.90
-0.06
0.85
0.13
0.03
0.85
-0.07
0.85
0.12
0.80
0.80
0.02
0.80
-0.08
0.75
0.11
0.75
0.01
0.75
-0.09
0.70
0.10
0.70
0.00
0.70
-0.10
0.65
0.09
0.65
-0.01
0.65
-0.11
0.60
0.08
0.60
-0.02
0.60
-0.12
0.55
0.07
0.55
-0.03
0.55
-0.13
0.50
0.06
0.50
-0.04
0.50
-0.14
0.45
0.05
0.45
-0.05
0.40
0.14
0.40
0.04
0.40
-0.06
0.35
0.13
0.35
0.03