AM1610 Precision Industrial PRT User's Guide

2 Specifications

2.1 Specifications
Temperature Range
Resistance at 0 °C
Temperature Coefficient
Accuracy
Drift
Short Term Stability
Thermal Shock
Hysteresis
Self-heating
Response Time
Measurement Current
Sensor Length
Sensor Location
Insulation Resistance
Sheath Material
Dimensions
External Leads
Handle Dimension
Handle Temperature Range* -50 °C to 160 °C
Optional Calibration
*Handle temperature outside this range will cause damage to the probe.
Rev. 201609V3
-60 °C to 160 °C
Nominal 100 Ω
0.00385 Ω/ Ω/°C
±0.05 °C over the full range
±0.04 °C at 0 °C after 100 hours at 160 °C
±0.01 °C
±0.01 °C after 10 times thermal cycles from minimum to
maximum temperatures
<=0.01 °C
50 mW/°C
5 seconds for 63% response to step change in water moving
at 3 feet per second
0.5 mA or 1 mA
30 mm
5 mm from tip
>1000 MΩ at room temperature
316 Stainless Steel
1610-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1610-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)
tm
Teflon
– insulated copper wire, 4 leads, 2.5 meters
1610-12: 15 mm (OD) X 65 mm (L)
1610-9: 10 mm (OD) X 50 mm (L)
NIST traceable calibration and data available per request:
1610-12: Ordering # 5017
4/10