AccuPulse NIBP Simulator Manuais de operação - Página 10

Procurar online ou descarregar pdf Manuais de operação para Equipamento de teste AccuPulse NIBP Simulator. AccuPulse NIBP Simulator 20 páginas. Hand held nibp simulator

User-Defined Mode Editing
This is part of an optional software upgrade
Figure UD3 shows the User-Def Edit menu for the BP Test Screen.
From the User-Def menu, pressing F3 (Edit) opens the current selection for editing and
brings up the User-Def Edit Menu.
F1 (Select) moves the highlight to select a parameter for editing.
Left/Right arrow keys also move the highlight to select a parameter for editing.
The Up/Down arrow keys change the value of the highlighted parameter.
F2 (Trim) opens the Trim Menu.
F4 (Enter) saves the changes.
Figure UD4 shows the User-Def Trim menu for the BP Test Screen.
From the User-Def Edit menu, pressing F2 (Trim) opens this additional menu level of
editing.
F1 (Select) chooses Delta Sys, Delta Dia, or the MAP label for editing.
Left/Right arrow keys also move the selection highlight.
The Up/Down arrow keys change the value of the highlighted parameter.
F2 (Trim) returns to the standard Edit Menu.
F4 (Enter) saves the changes.
User-Defined Mode Ranges and Use
This is part of an optional software upgrade
The User Defined parameters can be adjusted over a very broad range. For some
parameters the range well exceeds that of typical NIBP monitors currently available. This
extended range is provided for research and to explore the true absolute limits of an
instrument. We impose few limits on the combinations available and leave it to the user
to decide what is valid for their purpose or physiologically unrealistic.
General range limits:
Systolic range is 20 – 295 mmHg
Diastolic range is 10 – 260 mmHg
Pulse Rate range is 15 – 330 BPM
Amplitude range is 0 – 150 % of our nominal Adult setting
A/N (Adult/Neo) is a label only. The appropriate amplitude for neonatal use must be set
by adjusting the amplitude setting.
The actual value of MAP output is set internally by the rule of thirds and follows changes
of the SYS/DIA settings made by the user. By default this value label is not displayed. It
can be turned on using the Trim menu. The MAP value displayed is a label only. In the
Trim Menu a user can adjust the value displayed to match the expected monitor results,
but can not directly move the underlying MAP point.
When making adjustments a minimum pulse pressure (Sys – Dia) of 10 is enforced.
Attempting to move the systolic value below this spread will push the diastolic down as
well. Raising diastolic operates in similar fashion.
The maximum Trim value range is +/- 25 mmHg for both Delta Sys and Delta Dia. This
range is limited as needed when the Trim would otherwise meet the MAP point or the
end range of the AccuPulse Plus unit.
Fig. UD3.
All User Defined settings will be permanently stored until a user changes them. These
will be retained across a power cycle.
Seq Tab (Auto Sequence)
The User Defined option module also adds some extended functionality to Auto-
sequence while operating in the User-Def Mode. These changes allow the user to choose
any of the 45 User-Def points to define their own sequence.
Under the Seq Tab pressing F1 (Seq Config) opens a new screen and menu to select the
points to use.
User Defined Auto Sequence
F1 (Add) allows the user to add points to the sequence immediately following the current
highlight point.
Fig. UD4.
Up/Down arrow keys change the value of the highlighted sequence location.
Left/Right arrow keys move the highlight across the sequence list.
F2 (Del) will delete the currently highlighted point from the sequence.
F3 (Save) stores the currently displayed sequence for use.
F4 (Exit) leaves without saving any changes.
Note that points to run do not have to be adjacent in the User-Defined Mode list, nor in
any particular order, and a single point can be used more than once. (Example: 1, 18,
43, 44, 7, 32, 18, 32 )
All other Autosequence functions remain unchanged from the standard operation.
Mot Tab (Motion Artifact)
This is an optional software upgrade
The Motion Artifact option provides a way to mix noise into the normal pulse patterns.
While in BP Test mode F1 switches Motion On and Off.
Figure M1 shows the Motion Tab screen.
10
User-Defined mode feature extension
Sequence Configuration
Fig. M1