Accuris myVolt Touch Manual de instruções - Página 4
Procurar online ou descarregar pdf Manual de instruções para Fonte de alimentação Accuris myVolt Touch. Accuris myVolt Touch 6 páginas. 300v
2. Product Diagram
In constant amperage mode, the power supply will output the set
4 Pairs of
mAmp current parameter, and the voltage will fluctuate relative
Touch Screen
Electrical Outputs
to the resistance of the load (buffer concentration in the attached
Control and Display
electrophoresis tanks) up to a maximum of the voltage setting
that has been chosen.
FRONT
6.4 Automatic Cross Over
During an electrophoresis run, the power supply will attempt to
maintain the constant voltage or amperage setting, depending on
which parameter is set to "CONSTANT".
If, during a run, the non-constant parameter (limit) is met due to
the resistance of the load, the power supply will automatically
"cross over" so the non-constant parameter will now be the
constant and limiting parameter.
6.5 Maximum Power Limit
Folding
Recesses for
The myVolt Touch Power Supply has a maximum power
Front Feet
Stacking 2 Units
output of 80 Watts.
Per Ohms Law:
Power (P) = Current (I) x Voltage (V)
If the maximum power is reached during an electrophoresis
run, the non-constant parameter will be limited to prevent
.
exceeding the maximum power output
CAUTION:
It is not recommended to run the power supply
at or near the maximum wattage, as this will cause the power
supply and gel tank to get hot. Most electrophoresis buffer
concentration will not have such low resistance that the
Rating, Serial #
current and voltage level will come close to the maximum
AC Power Inlet
REAR
Label
power limit.
and Fuse
7
2