Siemens SIREMOBIL Compact L Manual de instruções de instalação e configuração - Página 8

Procurar online ou descarregar pdf Manual de instruções de instalação e configuração para Equipamento industrial Siemens SIREMOBIL Compact L. Siemens SIREMOBIL Compact L 28 páginas.
Também para Siemens SIREMOBIL Compact L: Instruções de instalação e configuração (4 páginas), Manual de instruções (21 páginas), Manual de instruções de instalação (6 páginas)

Siemens SIREMOBIL Compact L Manual de instruções de instalação e configuração
1 - 4

Safety information - risk of infection

WARNING

Protective conductor resistance test information

Observe the instructions in the safety rules for installation and repair
(ARTD-002.731.17 ...).
The protective conductor resistance must be measured after every intervention in the
system.
However, documentation of the measured values is required only during periodical
safety checks.
If parts / components that can decisively influence the protective conductor resistance are
replaced (e.g. replacement of the mains cable, replacement of the ON/OFF assembly,
replacement of multi-core connection cables, which also create the protective conductor
connection between parts of the system (e.g. monitor cable or C-arm cable)), or if
protective conductor connections have been repaired, then the protective conductor
resistance must be measured. The values must be documented and assessed in the
protective conductor resistance report.
NOTE
The measurement must be made according to DIN VDE 0751, Part 1 (see ARTD Part 2).
In this case the protective conductor resistance in the normal operating condition to all
conductive touchable parts of the system must be measured.
Make sure that control cables or data cables between the components of the system do
not imitate any protective conductor connection.
During the measurement the power cable and additional connection cables which also
create the protective conductor connection between parts of the system (e.g. monitor
cable between basic unit and monitor trolley) must be moved section by section to detect
broken conductors.
The protective conductor resistance may not exceed 0.2 ohms.
SIREMOBIL Compact L
System Manual
Danger of infections due to pathogens!
Non-compliance can lead to severe injuries and even death.
This product can be contaminated by infected blood or other
bodily fluids.
Avoid all contact with blood or other bodily fluids!
Strictly observe the safety information in ARTD-002.731.37...
regarding prevention of infectious diseases during customer
service calls.
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding
maintenance procedures or safety checks.
A sudden or unexpected increase of the measured values,
even if the limit value of 0.2 ohms is not exceeded,
indicates errors in the protective conductor connections.
(Protective conductor or contacts).
n.a.
SPR2-130.841.02
Rev. 01
Page 4 of 6
07.05
CS PS 24
General
1
1
Siemens AG
Medical Solutions