Safety Inspection
3-
3
Safety Inspection
3.1
Mechanical safety
SIM
Cover panels
•
Remove the covers from the SIREMOBIL stand and monitor trolley.
•
Check cover panels for damage.
SIM
Cable deflectors
•
Inspect the cable deflectors on the stand and on the monitor trolley.
SIM
Foot brake
•
Test the function of the foot brakes on the SIREMOBIL stand and monitor trolley on a
level floor.
SIM
Brakes
•
Use the spring balance to test whether the orbital, angulation, swivel and horizontal
movement brakes apply the prescribed force. (Refer to service instructions SIREMO-
BIL ISO-C)
SIM
C-arm
•
Check the C-arm for damage to the carbon fiber structures (visual inspection). If there is
damage, proceed as described in the warning in chapter 1.
•
Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
•
Orbital movement: Once the orbital brake is released, the C-arm can be moved only in a
limited manner. Check whether the orbital movement is smooth.
SIM
Wheels
•
Move the SIREMOBIL in a straight line on a level floor.
•
Evaluate the straight and quiet movement of the SIREMOBIL.
•
Replace any defective wheels or castors.
•
Check the steering function.
SIM
Lifting column
•
Switch the system on
•
Electrically raise and lower the lifting column,
- while listening for noises and checking for play in the bearings.
- The lifting column movement must switch off automatically when the end positions
are reached.
•
Switch off the system.
•
Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIM
Emergency Stop switch
•
Press the Emergency Stop switch. The lifting column may not be raised or lowered.
•
Press the Emergency Stop switch again. The lifting column can be moved again.
Siemens AG
Medical Solutions
SPR2-230.101.02.06.02
07.05
Page 29 of 46
CS PS 24
29
SIREMOBIL Iso-C