Operating (hardware)

3.3 Verificator software - sensor - Insulation

Sensor:
Press GO to continue
3.3
Verificator software - sensor - Insulation
If a SENSORPROM
The bar indicates how the test is progressing.
Press ESC to cancel the test.
T2:3 Insulation test
0%- ##
Press ESC to cancel
Insulation test finished
Press ESC to cancel further tests.
T2:3 Insulation test
0% -###############
Press ESC to cancel
Is only displayed if the insulation resistance of the sensor/system is insufficient. See "On site
trouble shooting".
Value shows the measured value in nV.
Press ESC to quit.
E2: Sensor failed
Insulation fault
Value xxxxxx xxx
Press ESC to quit
Sensor/system passed the insulation test
Verification continues automatically.
I2: Sensor Passed
Insulation OK
Please wait ....
12
No SENSORPROM
unit is install, then the verification test will continue with the insulation test.
®
-100%
-100%