5.3
Unit with remote-located LVDT conditioner card
LVDT to SF500 Integrator
*Where separation between the integrator and LVDT conditioner exceeds 150 m (500 ft):
1. Remove the jumpers from SF500 terminals 11/12 and 13/14
2. Run additional conductors:
– from SF500 terminal 12 to conditioner terminal block marked "Integrator EXC+"
– from SF500 terminal 13 to conditioner terminal block marked "Integrator EXC-"
For further connection information on specific LVDTs, consult Siemens.
** For Encapsulated LVDT (hazardous)
• YEL = WHITE
• BLU = ORANGE
SITRANS WFS300
Operating Instructions, 03/2021, A5E32880243-AC
5.3 Unit with remote-located LVDT conditioner card
• BLU = ORANGE
• GRN = YELLOW
• BLK = Not connected
Connecting
19