Siemens ULTRAMAT 6F Manual de Instruções - Página 11
Procurar online ou descarregar pdf Manual de Instruções para Equipamento industrial Siemens ULTRAMAT 6F. Siemens ULTRAMAT 6F 49 páginas. Gas analyzer for the determination of ir-absorbing gases
Também para Siemens ULTRAMAT 6F: Manual de equipamento adicional (20 páginas), Manual de Segurança (24 páginas)
3.1.2.3
Dimensions for preparing installation
Figure 3-1
CALOMAT 6F, CALOMAT 62F, OXYMAT 6F, and ULTRAMAT 6F:
dimensions for preparing installation, front view and side view
3.1.2.4
Connection of the flame arrestors
Application
Sample gases that are occasionally explosive may only be connected to the CALOMAT 6F, CALOMAT 62F and
OXYMAT 6F gas analyzers if the sample gas inlet and outlet have each been provided with a flame arrestor with adapter.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018
11