Siemens UROSKOP D Instruções adicionais de instalação, operação e manutenção - Página 22
Procurar online ou descarregar pdf Instruções adicionais de instalação, operação e manutenção para Equipamento industrial Siemens UROSKOP D. Siemens UROSKOP D 26 páginas.
2 - 14
2
Fig. 6
2.8
Function inspection
PMF
Spotfilm device
•
Load a 24 x 18 cm cassette.
•
Set R or L exposure (1/Fig. 6).
•
Set a random time on the timer (2/Fig. 6).
•
Release an exposure with 45 kV and automatic exposure control.
•
Check the film exposure.
The R or L as well as the time indication must be clearly visible on the developed film.
PMF
Cassette recognition
•
Load a 43 x 35 cassette horizontally.
The cassette must be retracted into the spotfilm device and then ejected back into the
loading position.
The spotfilm device should sound an acoustic warning.
NOTICE
UROSKOP D
Inspection and maintenance
1
The 43 cm horizontal format has not been included in the cassette
program.
RLL5-310.101.02
Rev. 02
01.97
Page 14 of 16
TDSP 1
Medical Engineering
Siemens AG