General information
1
General information
NOTICE
1.1
Required documentation
depending on the system)
•
UROSKOP D1 circuit diagram
•
UROSKOP D2 circuit diagram
•
UROSKOP D3 circuit diagram
•
Adjustment instructions
•
Uroskop D1/D2 start-up instructions
•
Uroskop D3 start-up instructions
•
Maintenance protocol
1.2
Required tools, measurement and auxiliary equipment
•
Protective conductor meter
•
DVM e.g. Fluke 8060
•
Torque wrench, to 100 Nm
•
Spring scale, to 2.5 N
•
Spring scale, to 100 N
•
Standard service equipment
1.3
Required lubricant
•
Grease gun for Longtime PD2
•
Longtime PD2 - grease
•
Optimol Optiflex ELM paste
•
Silicon paste
Siemens AG
Medical Engineering
The sequence for complete inspection and maintenance is
described on the following pages.
The safety inspection may be performed alone; however, the
operating values check must be performed in conjunction with
the safety inspection.
20 g tube
1 kg can
1 kg can
100 g tube
RLL5-310.101.02
Rev. 02
01.97
G5353
G5354
G5403
RLL5-310.071.01.02.01
RLL5-310.034.01.04.01
RLL5-310.034.02.03.01
RLL5-310.105.02.02.01
44 15 899 RV090
84 27 684 RE999
71 09 010
34 91 271
73 95 445
15 90 293
20 49 914
Page 1 of 4
TDSP 1
1 - 1
UROSKOP D